Table 1.

Nitric oxide and oxygen flow (in liters/minute) were set to obtain a desired NO inspiratory concentration (50, 150, and 250 ppm) at a desired FiO2. We set medical air flow at 15 Liters/min in every setting. We used two different tanks with different NO concentrations (800 and 857 ppm).

NO/N2tank concentration = 800 ppm
NO (ppm) NO2 (ppm) NO flow (L/min) O2 flow (L/min)
FiO2 = 0.21 49 0.13 1 0.36
145.5 0.72 3.46 1.11
249.5 1.57 6.99 2.10
FiO2 = 0.30 48.5 0.18 1.2 3.50
140 0.97 4.25 4.82
250 2.35 9.10 7.30
FiO2 = 0.41 48 0.18 1.40 7.12
145 1.17 5.08 10.20
241.5
2.61
11.91
17.40
NO/N2 tank concentration = 857 ppm

NO (ppm)
NO2 (ppm)
NO flow (L/min)
O2 flow (L/min)
FiO2 = 0.21 47.50 0.24 0.92 0.58
143.50 0.77 3.07 1.16
237.50 1.98 6.18 2.37
FiO2 = 0.30 49 0.26 1.05 3.46
143 1.24 3.84 5.40
253 2.79 7.66 7.22
FiO2 = 0.42 40 0.29 1.22 8.48
142 1.40 4.48 10.74
236 3.11 10.05 15.70