Supplemental Data Table S2.

Analyte and internal standard MS parameters

Analyte Transition MRM transition (m/z) Collision energy (V) Cell accelerator voltage (V) RT (min)
Melatonin Quantifier 233.1 → 174.1 9 4 1.3
Qualifier 233.1 → 159.2 33 4 1.3

Melatonin-d4 Quantifier 237.1 → 178.1 17 4 1.3

Cortisol Quantifier 363.1 → 120.9 25 4 2.2
Qualifier 363.1 → 327.4 13 4 2.2

Cortisol-d4 Quantifier 367.4 → 121.1 25 4 2.2

Abbreviations: MS, mass spectrometry; MRM, multiple reaction monitoring; RT, retention time.