Skip to main content
. 2021 Jan 1;41(1):68–76. doi: 10.3343/alm.2021.41.1.68

Supplemental Data Table S6.

Correlation analysis between seven routine assays and the means of all assays for 85 clinical samples

Assay ALT AST


r2 Slope (95% CI) Intercept (95% CI) Sy.x r2 Slope (95% CI) Intercept (95% CI) Sy.x
Biosino 0.9999 1.04 (1.04–1.05) −0.88 (−1.11 to −0.66) 0.63 0.9999 1.02 (1.02–1.02) −0.30 (−0.52 to −0.08) 0.69

BSBE 0.9999 1.07 (1.07–1.08) −0.58 (−0.87 to −0.28) 0.83 0.9998 0.94 (0.94–0.95) 0.40 (0.06 to 0.74) 1.06

Diasys 0.9999 0.98 (0.98–0.98) 1.01 (0.76 to 1.26) 0.71 0.9999 1.02 (1.01–1.02) 0.39 (0.14 to 0.63) 0.76

KHB 0.9999 0.95 (0.94–0.95) −0.95 (−1.22 to −0.69) 0.75 1.0000 0.96 (0.96–0.96) −0.34 (−0.51 to −0.17) 0.52

LEADMAN* 0.9998 0.96 (0.96–0.97) 0.14 (−0.22 to 0.50) 1.02 0.9997 1.06 (1.06–1.07) 0.75 (0.26 to 1.24) 1.51

Maccura 0.9998 1.00 (1.00–1.00) 0.80 (0.43 to 1.17) 1.04 0.9999 1.00 (1.00–1.01) 0.09 (−0.21 to 0.40) 0.93
*

LEADMAN routine assay.

Abbreviations: r2, determination of coefficient; CI, confidence interval; Sy.x, residual standard deviation.