TABLE 1.

Analytical limits of detection for SARS-CoV-2

Assay No. of SARS-CoV-2 detected/no. tested at a viral concn (cp/ml) of:
Exp. LODa Probit LOD
3,160 1,000 316 100 32 10
Abbott m2000 ND ND 20/20 19/20 20/20 16/20 32 53
Roche Cobas ND ND 20/20 20/20 20/20 20/20 ≤10 ≤10
Panther Fusion 5/5 5/5 5/5 20/20 15/20 1/20 100 74
DiaSorin Simplexa 5/5 5/5 20/20 19/20 14/20 ND 100 167
GenMark ePlex 5/5 5/5 5/5 19/20 8/20 ND 100 190
Abbott ID NOW 5/5 5/5 24/25 14/25 9/20 0/5 316 511
CDC ABI 7500 (MP24) 5/5 5/5 20/20 15/20 1/5 ND 316 237
CDC LC 480 (MP24) 5/5 5/5 19/20 14/20 1/5 ND 316 499
CDC ABI 7500 (EZ1) 5/5 5/5 20/20 20/20 13/20 ND 100 85
CDC LC 480 (EZ1) 5/5 5/5 20/20 19/20 10/20 ND 100 185
a

Exp. LOD, experimental limit of detection; Probit LOD, Probit limit of detection; ND, not done; ABI 7500, Applied Biosystems 7500 Fast real-time PCR system; LC 480, Roche LightCycler 480 II real-time PCR system; MP24, Roche MagaPure 24 extractor; EZ1, Qiagen EZ1 extractor.