Skip to main content
. 2020 Aug 24;58(9):e01535-20. doi: 10.1128/JCM.01535-20

TABLE 2.

Sample volumes tested for SARS-CoV-2

Assay Minimum input vol (μl) Sample processed (μl) Elution vol (μl) Eluate added to reaction (μl) Effective sample vol(s) tested (μl)
Abbott m2000 760 500 80 40 250
Roche Cobas 600 400 50 50 400
Panther Fusion 500a 360 50 5 25.4
DiaSorin Simplexa b 50
GenMark ePlex 200
Abbott ID NOW 100 (target), 100 (internal control)
CDC ABI 7500 (MP24) 350 200 100 5 10
CDC LC 480 (MP24) 350 200 100 5 10
CDC ABI 7500 (EZ1) 400 400 90 5 22.2
CDC LC 480 (EZ1) 400 400 90 5 22.2
a

A 500-μl sample + 710-μl lysis buffer = 1,210 μl input volume.

b

Sample-to-answer tests.