Table 3.

Precision and accuracy of LVFX in rat liver tissues, urine, and feces (n = 5).

Biosamples Spiked concentration (ng/mL) Intraday Interday
Calculated concentration (ng/mL) Precision (RSD, %) Accuracy (%) Calculated concentration (ng/mL) Precision (RSD, %) Accuracy (%)
LVFX in liver tissue 10.02 9.33 ± 1.03 11.03 93.07 9.18 ± 0.55 5.94 91.59
100.20 104.61 ± 3.89 3.72 104.40 103.86 ± 4.09 3.94 103.65
5010.00 4815.20 ± 158.68 3.30 96.11 4803.63 ± 111.96 2.33 95.88

LVFX in urine 10.02 9.49 ± 0.85 8.99 94.67 10.46 ± 0.51 4.85 104.35
200.40 209.45 ± 10.49 5.01 104.52 207.55 ± 12.98 6.25 103.57
4008.00 4183.48 ± 102.27 2.44 104.39 3875.60 ± 97.53 2.52 96.70

LVFX in feces 10.02 9.54 ± 0.86 9.03 95.17 10.54 ± 0.73 6.90 105.16
20.04 20.56 ± 1.16 5.66 102.58 19.54 ± 0.91 4.64 97.53
200.40 215.78 ± 13.68 6.34 107.67 190.05 ± 20.49 10.78 94.83