Recovery of Cu, Fe, Zn, and Cd and interfering elements for Cu obtained from applying the trace metal separation protocol to whole blood and plasma samples

Fraction Elements Whole blood (n = 3) Seronorm™ Trace Elements L-3 Plasma (n = 10) From 10 donors
AVE ± SD AVE ± SD
Cu Cu 99.5 ± 0.1% 99.5 ± 0.1%
Na 0.1 ± 0.1% 0.3 ± 0.3%
Mg 0.4 ± 0.3% 1.4 ± 1.8%
Ca 1.6 ± 0.4% 1.9 ± 1.3%
P 0.1 ± 0.1% 0.6 ± 0.8%
S 0.3 ± 0.5% 0.3 ± 0.5%
Fe Fe 99.7 ± 0.2% 98.1 ± 0.6%
Zn Zn 99.6 ± 0.1% 98.6 ± 0.7%
Cd Cd 93.8 ± 3.5% -