Skip to main content
. 2020 Aug 17;34(8):868–873. doi: 10.1089/end.2020.0097

Table 2.

Average Mass Changes of All Urinary Tract Device Samples After a 90-Day Trial

Material Average initial mass (mg) Average final mass (mg) Average percent change in mass
Stent 138 231 63.78%
Uncoated Pellethane® 134 149 11.50%
HEMA-Coated Pellethane 140 144.5 2.90%
TETRA-Coated Pellethane 142 143 0.60%

HEMA = 2-hydroxyethyl methacrylate; TETRA = tetraethylene glycol dimethyl ether.