. 2020 Aug 17;34(8):868–873. doi: 10.1089/end.2020.0097

Table 3.

Presence of Magnesium, Calcium, and Phosphorous Content Detected by Inductively Coupled Plasma Mass Spectrometry

Material Magnesium*
Calcium*
Phosphorus**
ppb SD %RSD ppb SD %RSD ppb SD %RSD
Stent Set 1 125.57 3.70 2.95 564.09 10.39 1.84 417.70 2.62 0.63
Set 2 41.33 1.31 3.18 215.70 4.63 2.15 545.37 3.24 0.59
Pellethane Set 1 36.85 0.94 2.56 242.06 5.59 2.31 193.91 1.59 0.82
Set 2 17.38 0.60 3.44 168.42 8.02 4.76 8.70 1.32 15.23
HEMA-Coated Pellethane Set 1 7.79 0.37 4.74 87.68 4.74 5.41 84.33 1.30 1.54
Set 2 7.55 0.55 7.32 43.08 2.95 6.84 292.55 1.95 0.67
TETRA-Coated Pellethane Set 1 33.14 1.81 5.45 185.38 4.01 2.16 360.69 2.28 0.63
Set 2 33.53 1.04 3.11 271.80 5.26 1.94 367.47 17.68 4.81
*

Isotopes measured in low resolution (R = 300): 26Mg, 43Ca, 72Ge (internal standard)

**

Isotopes measured in medium resolution (R = 2500): 31P.

%RSD = % relative standard deviation; SD = standard deviation.