Skip to main content
. 2020 Aug 8;25(16):3608. doi: 10.3390/molecules25163608

Table 1.

Results of Intra-day and inter-day accuracy and reproducibility assays. The quality controls concentrations are respectively 1, 3, 400, 650 µg/L for LLOQ, QC low, QC medium and QC high.

INTER-DAY
CBD THC
Dev. Std. (σ) CV % Accuracy % Dev. Std. (σ) CV % Accuracy %
LLOQ 0.01 6% −13% 0.01 5% −12%
QC low 0.22 10% 14% 0.32 9% 12%
QC medium 0.57 14% −4% 0.52 14% −7%
QC high 0.48 5% −5% 0.50 5% −1%
INTRA-DAY
CBD THC
Dev. Std. (σ) CV % Accuracy % Dev. Std. (σ) CV % Accuracy %
LLOQ 0.01 6% 14% 0.03 14% 13%
QC low 0.02 11% 9% 0.08 4% 6%
QC medium 0.07 6% 12% 0.11 10% 11%
QC high 1.48 13% 6% 0.40 4% 5%