Table 2.

Bald-Altman comparison among the participants for all the four parameters

Parameter Device Mean±SD Bald-Altman comparison
MD 95% CI
SpO2 (%) Philips Intellivue MX 500 96.96±2.09 Reference
ECIL device 97.09±2.27 −0.12 −2.36-2.11
Temperature (°F) Philips Intellivue MX 500 97.67±0.79 Reference
ECIL device 97.42±0.80 0.25 −0.25-0.75
Respiratory rate per minute Philips Intellivue MX 500 18.44±4.05 Reference
ECIL device 17.95±3.93 0.49 −2.05-3.04
Pulse rate per minute Philips Intellivue MX 500 80.87±10.7 Reference
ECIL device 80.60±10.61 0.27 −2.69-3.32

ECIL: Electronics Corporation of India Limited, SD: Standard deviation, CI: Confidence interval, MD: Mean difference