Table 2.
Parameter | Device | Mean±SD | Bald-Altman comparison |
|
---|---|---|---|---|
MD | 95% CI | |||
SpO2 (%) | Philips Intellivue MX 500 | 96.96±2.09 | Reference | |
ECIL device | 97.09±2.27 | −0.12 | −2.36-2.11 | |
Temperature (°F) | Philips Intellivue MX 500 | 97.67±0.79 | Reference | |
ECIL device | 97.42±0.80 | 0.25 | −0.25-0.75 | |
Respiratory rate per minute | Philips Intellivue MX 500 | 18.44±4.05 | Reference | |
ECIL device | 17.95±3.93 | 0.49 | −2.05-3.04 | |
Pulse rate per minute | Philips Intellivue MX 500 | 80.87±10.7 | Reference | |
ECIL device | 80.60±10.61 | 0.27 | −2.69-3.32 |
ECIL: Electronics Corporation of India Limited, SD: Standard deviation, CI: Confidence interval, MD: Mean difference