Skip to main content
. 2020 Jul 29;12(8):2271. doi: 10.3390/nu12082271

Table 3.

Inter-day precision (n) of vitamin D metabolites in plasma, breastmilk, and infant formula samples measured by (a) Waters Quattro PremierTM/XE and (b) Agilent 6495B.

Analyte
a. Waters Quattro PremierTM/XE
Plasma (n = 10) 25(OH)D2 25(OH)D3 3-Epi25(OH)D3
Medium1
 Average (nmol/L) - 55.5 ± 5.2 -
 RSD (%) 9
b. Agilent 6495B
Plasma (n = 10) 25(OH)D2 25(OH)D3 3-Epi25(OH)D3
Low2
 Average (nmol/L) 30.2 ± 2.7 42.9 ± 3.4 34.5 ± 2.2
 RSD (%) 9 8 6
Medium2
 Average (nmol/L) 71.2 ± 4.2 102.6 ± 5.6 48.2 ± 3.6
 RSD (%) 6 5 7
High2
 Average (nmol/L) 115.3 ± 8.9 162.7 ± 9.0 99.9 ± 5.6
 RSD (%) 8 6 6
Breastmilk (n = 10) 25(OH)D2 25(OH)D3 3-Epi25(OH)D3
Low3
 Average (nmol/L) <7 12.0 ± 2 5.8 ± 1.7
 RSD (%) 17 30
Medium3
 Average (nmol/L) 39.9 ± 2.7 65.0 ± 5.2 45.7 ± 2.6
 RSD (%) 7 8 6
High3
 Average (nmol/L) 97.9 ± 4.6 114.8 ± 8 92.0 ± 9
 RSD (%) 5 7 10
Infant formula (n = 10) 25(OH)D2 25(OH)D3 3-Epi25(OH)D3
Low4
 Average (nmol/L) <7 <4.3 <1.7
 SD
 RSD (%)
Medium4
 Average (nmol/L) 21.4 ± 2.3 66.6 ± 2.7 32.2 ± 1.8
 RSD (%) 11 4 6
High4
 Average (nmol/L) 93.2 ± 8.7 112.1 ± 4.1 107.1 ± 3.6
 RSD (%) 9 4 3