Skip to main content
. Author manuscript; available in PMC: 2021 Sep 1.
Published in final edited form as: ASAIO J. 2020 Sep-Oct;66(9):1016–1024. doi: 10.1097/MAT.0000000000001145

Table 2 –

Mitral valve parameters before and after LV thinning

Before After p-value
MR jet area/ Left atrial area (%) 0.00 24.41 ± 16.59 0.0080
Anterior leaflet tenting area (mm2) 87.77 ± 15.44 290.9 ± 78.56 0.0004
Posterior leaflet tenting area (mm2) 42.93 ± 23.66 119.0 ± 61.24 0.0160
Systolic anterior angle (°) 32.6 ± 9.5 48.5 ± 13.0 0.0033
Systolic posterior angle (°) 43.4± 5.0 50.2± 12.6 0.2394
Anterior leaflet tethering angle (°) 159.0 ± 10.8 137.2 ± 10.5 0.0002
Posterior leaflet tethering angle (°) 122.0 ± 16.6 114.1 ± 13.7 0.1311