Table 4.

Summary of active materials, sensing principles used in wearable sensors along with their sensing properties.

SN Active Material
(Structure)
Sensing
Principles
Limit of
Detection
(kPa) 1
Maximum
Detection
(kPa) 2
Sensitivity
(kPa−1) 3
Reference Notes
1.0 CNT/PDMS
(Porous Structure)
Piezoresistive 0.25 100.0 0.588 4 [257]
  • No blood pressure measurement application.

  • Not all properties were reported (e.g., hysteresis, stability test).

  • High limit of detection.

2.0 CNT@EcoFlex
(Buckled Structure)
Piezocapacitive 50.0 1000 230 [244]
  • No blood pressure measurement application.

  • Not all properties were reported (e.g., hysteresis, stability test).

  • High limit of detection.

3.0 ACNT-Graphene
/PDMS
(CVD)
Piezoresistive 0.0006 0.3 19.8 [259]
  • No blood pressure measurement application.

  • Response time is <16.7 ms. 5

  • Stability for more than 35,000 loading/unloading cycles. 6

  • Operating voltage is 0.03 V. 7

  • Low maximum detection.

4.0 CNT/PDMS
(Patterned Microstructure)
Piezoresistive 0.0002 59.0 15.1 [258]
  • No blood pressure measurement application.

  • Performance changed under cycle-test with constant pressure.

  • Not all properties were reported.

  • The operating voltage is 10.0 V. 7

5.0 VACNT/PDMS
(T-CVD)
Piezoresistive 0.002 10.0 0.3 and
up to 0.7
[260]
  • Blood pressure measurement application without calibration, need to assess accuracy.

  • Response time is around 162 ms. 5 Acceptable reproducibility over 5000 cycles of pressure loading/ unloading. 6

6.0 SWCNTs/PDMS
(Silk Molded Microstructure)
Piezoresistive 0.0006 1.2 1.8 [272]
  • No blood pressure measurement application but heart pulse.

  • Response time is 10 ms. 5

  • Stability for around 67,500 loading/unloading cycles. 6

  • Operating voltage is 2.0 V. 7

7.0 CB@PU Sponge Piezoresistive 0.091 16.4 0.068 [240]
  • No blood pressure measurement application but heart pulse.

  • Stability for around 50,000 loading/unloading cycles. 6

  • Response time is 20 ms. 5

8.0 MWCNT-rGO@PU Foam Piezoresistive 0.0035 2.7 0.022 [256]
  • No blood pressure measurement application but heart pulse.

  • Stability for around 5000 loading/unloading cycles. 6

  • Response time is < 30 ms. 5

9.0 SWCNTs/PDMS Optical 1.0 - 0.2 [201]
  • No blood pressure measurement application.

  • Response time is 300 ms. 5

  • No stability test information was reported.

  • Transparent optical application.

10.0 CNT/3D Microporous Elastomeric Dielectric Layer Piezocapacitive 0.0001 130 0.601 [194]
  • No blood pressure measurement application but heart pulse.

  • Stability for around 1000 loading/unloading cycles. 6

  • Operating voltage can be 1.0 V. 7

11.0 Graphene@PU Piezoresistive 0.009 10.0 0.26 [273]
  • No blood pressure measurement application but heart pulse.

  • Stability for around 10,000 loading/unloading cycles. 6

  • Not all properties were reported.

12.0 Graphene
(MEMS)
Piezoresistive 0.1 - 3.4 × 10−6 [274]
  • No blood pressure measurement application.

  • Response Time is 0.9 s. 5

  • Operating voltage is 2.5 V. 7

  • Not all properties were reported.

13.0 Graphene Paper Piezocapacitive 2.0 20.0 17.2 [275]
  • No blood pressure measurement application, but heart pulse.

  • Response time is120 ms. 5

  • Stability for around 300 loading/unloading cycles. 6

  • Fabricated by Dip Coating.

14.0 Graphene Electrode
(T-CVD)
Piezocapacitive 4.4 × 10−5
(1 mg)
- 3.19 [247]
  • No blood pressure measurement application, but it sounds to be suitable for this application.

  • The applied voltage is 3.3 V. 7

  • Stability for around 500 loading/unloading cycles. 6

  • Response time is 30 ms. 5

15.0 Graphene
(Porous GS)
Piezoresistive 0.3 10.0 0.046 [276]
  • No blood pressure measurement application but heart pulse.

  • Stability for around 200 loading/unloading cycles. 6

16.0 Graphene Electrode Piezocapacitive 8.0 - 6.55 [268]
  • No blood pressure measurement application.

  • Stability for around 500 loading/unloading cycles. 6

  • Response time is 70ms. 5 =

17.0 Suspended
Graphene /Polymer
(Heterostructure
Membranes)
Piezocapacitive 80.0 - 123ZF [277]
  • No blood pressure measurement application.

  • Not all properties were reported.

18.0 Graphene Tribotronics FET 1.0 - 0.02 [278]
  • No blood pressure measurement application.

  • Stability for around 1700 loading/unloading cycles. 6 =

  • Response time is 30 ms. 5 =

  • Operating voltage can be 0.5 V. 7 =

  • Suitable for electronic skin and touch screen applications.

19.0 rGO/PANI
Wrapped Sponge
Piezoresistive 0.1 27.0 0.152 [279]
  • No blood pressure measurement application.

  • Response time is around 96 ms. 5 =

  • Stability for around 9000 loading/unloading cycles. 6 =

  • High current output (i.e., ∼300 μA at 1.0 V. 7 =

20.0 PNIPAm/CMC/ rGO DN Hydrogel Thermo-resisitive - 800.0 - [280]
  • No blood pressure measurement application.

  • It is also sensitive to temperature (30 °C–45 °C).

21.0 rGO/PU
Sponge
Piezoresistive 4.84 × 10−5 (1.1 mg) - 0.21 [281]
  • No blood pressure measurement application.

  • Stability for around 10,000 loading/unloading cycles. 6 =

  • Response time is about 100 ms. 5 =

  • Not all properties were reported.

22.0 rGO/PDMS Film
(Pattered Micropyramid)
Piezoresistive 0.0015 1.4 5.5 [243]
  • No blood pressure measurement application.

  • Response time is 0.2 ms. 5 =

  • Stability for around 5000 loading/unloading cycles at 100 Pa. 6 =

  • The applied voltage is 1.0 V. 7 =

23.0 rGO Films with Continuous
Gradient
Wrinkles
Piezoresistive 0.0042 3.0 178.0 [282]
  • No blood pressure measurement application.

  • Stability for around 1200 loading/unloading cycles. 6 =

  • Response time is 131 ms. 5 =

  • Operating Voltage can be 1.0 V. 7 =

24.0 Large-Scale
Polystyrene
Ball@rGO Core Shell NPs
Piezoresistive 0.003 3.0 50.9 [283]
  • No blood pressure measurement application but heart pulse.

  • Stability for around 20,000 loading/unloading cycles. 6

  • Response time is 50 ms. 5

  • Operating Voltage can be 1.0 V. 7

25.0 Graphene
(Electrode Microconformal)
Piezocapacitive 4.4 × 10−5
(1 mg)
- 7.68 [190]
  • No blood pressure measurement application but heart pulse.

  • Stability for around 5000 loading/unloading cycles. 6

  • The applied voltage is 1.0 mV. 7

  • Response time is 30ms. 5

26.0 Integrated
Arrays of Air-Dielectric Graphene Transistors
FET 0.25 3000 2.05 × 10−4 [270]
  • No blood pressure measurement application.

  • Response Time is 30 ms. 5

  • Stability for around 1000 loading/unloading cycles. 6

  • The gate voltage is 25.0 V. 7

27.0 Graphene Transistor
Array
(Direct-Contact Tribotronic Planar)
FET 0.16 mm−1 - - [269]
  • No blood pressure measurement application.

  • Stability for around 1000 loading/unloading cycles. 6

  • Response time is 15 ms. 5

  • Gate voltage can be 2.0 V. 7

28.0 Graphene
(Direct Laser Scribing PDMS)
Piezoresistive 0.028 - 480.0 [271]
  • No blood pressure measurement application.

  • Stability for around 4000 loading/unloading cycles. 6

  • Response time is 0.002 ms. 5

  • Operating voltage can be 5.0 V. 7

29.0 3D Graphene
Film
(Fingerprint Like Patterned)
Piezoresistive 0.0002 75.0 110.0 [242]
  • No blood pressure measurement application but heart pulse.

  • Stability for around 10,000 loading/unloading cycles. 6

  • Response time is 30 ms. 5

  • Operating voltage can be 3.0 V. 7

30.0 GO (Spray Coating through a Stencil Mask) Piezocapacitive 0.24 × 10−3 - 0.8 [192]
  • No blood pressure measurement application.

  • Response time is around 100 ms. 5

  • Maximum hysteresis is 5.0%@ 1.4 kPa.

  • Stability for around 100 loading /unloading cycles. 6

31.0 MoS2/GPN /Ecoflex
(T-CVD)
Piezoresistive 0.6 25.4 6.06 [284]
  • Suitable for blood pressure measurement application.

  • Stability for around 4000 loading/unloading cycles. 6

  • Response Time is 200 ms. 5

32.0 PVA NWs/ Wrinkled
Graphene Film
Piezoresistive 0.00224 - 28.34 [285]
  • No blood pressure measurement application.

  • Stability for around 6000 loading/unloading cycles. 6

  • The applied voltage is 0.1V. 7

  • Not all properties were reported.

33.0 rGO Film/
PDMS Arrays
Piezoresistive 0.0013 225.0 1.71 [238]
  • No blood pressure measurement application.

  • Stability for around 5000 loading/unloading cycles. 6

  • Response time is 6 ms. 5

  • Operating Voltage can be 1.0 V. 7

34.0 P(VDF-TrFe)
/rGO
Piezoresistive 0.0012 - 15.6 [251]
  • No blood pressure measurement application.

  • Stability for around 100,000 loading/unloading cycles. 6

  • Response time is 5 ms. 5

  • The applied voltage is 1.0 mV. 7

35.0 PEDOT:PSS
/PUD
(Pattered Micropyramid)
Piezoresistive 0.023 8.0 10.3 [176]
  • No blood pressure measurement application but heart pulse.

  • Response time 0.2 s. 5

  • Stability 800 loading/unloading cycles. 6

  • Applied Voltage is 0.2 V. 7

36.0 PVDF-HFP/PEDOT
(3D Electrospun Nanofibers)
Piezoresistive 0.001 30.0 13.5 [286]
  • No blood pressure measurement application but heart pulse.

  • Stability for around 10,000 loading/unloading cycles. 6

  • Response time is around 0.4s. 5

  • The applied voltage is 1.0 V. 7

37.0 P(VDF-TrFe) (Electrospun Nanofiber) Piezoresistive 0.0001 0.012 0.00041 [287]
  • No blood pressure measurement application.

  • Stability for around 10,000 loading / unloading cycles. 6

38.0 [PPy@PVA-co-PE] and POE Nanofibers Piezoresistive 0.0013 7.0 1.24 [288]
  • Suitable for blood pressure measurement application.

  • Stability for around 250 loading/unloading cycles. 6

  • Operating voltage can be 2.0 V. 7

39.0 Au NWs/Tissue Paper
(Dip-Coating)
Piezoresistive 0.013 - 1.14 [289]
  • No blood pressure measurement application but heart pulse.

  • The operating voltage is 1.5 V 7 with low energy consumption that is <30 µW.

  • Response time is <17 ms. 5

  • High Stability for around 50,000 loading/unloading cycles.

40.0 Au NP Densely Packed µNW based Pressure Piezoresistive 0.025 0.0801 [290]
  • No blood pressure measurement application.

  • Stability for around 6000 loading/unloading cycles. 6

  • Response time is <270 ms. 5

  • Not all properties were reported.

41.0 Ag IDEs and PdOx NP
(Percolative Metal NP Arrays)
Piezoresistive 0.0005 1.0 0.13 [248]
  • No blood pressure measurement application.

  • Stability for around 500 loading/unloading cycles. 6

  • Hysteresis induced by a 1.0 kPa applied pressure showed a shift of 0.012% from the initial value.

42.0 AG NWs
(Ag NW Flower)
Piezocapacitive 0.0006 115.0 1.54 [195]
  • No blood pressure measurement application.

  • Stability for around 5000 loading/unloading cycles. 6

  • Not all properties were reported.

43.0 Ge/Si Core/shell NW PSR
(OLED)
FET 1.0 - 42.7 [210]
  • No blood pressure measurement application.

  • Response time is 1 ms.5

  • Operating voltage is 10 V. 7

  • Not all properties were reported.

44.0 MOF
(CuTCNQ)
Piezoresistive 0.00073 3.0 6.25 [249]
  • Suitable for blood pressure measurement application.

  • Stability for around 10,000 loading/unloading cycles. 6

  • Response time is 10 ms. 5

  • Operating voltage can be 1.0 V. 7

  • MOF is relatively expensive.

45.0 C-MOF/PANIF @PU Sponge Piezoresistive 0.001 60.0 158.26 [291]
  • No blood pressure measurement application, but heart pulse.

  • Stability for around 15,000 loading/unloading cycles. 6

  • Response time is < 22 ms. 5

46.0 C-MOF-5
Derived Porous Carbon
Piezoresistive 0.02 1.0 15.63 [292]
  • No blood pressure measurement application.

  • Stability for around 2000 loading/unloading cycles. 6

  • Response time is < 65 ms. 5

  • Operating voltage is 3.0 V. 7

47.0 MXenes
Nanosheets (Ti3C2Tx)
Piezoresistive 0.0102 30.0 0.55 [200]
  • No blood pressure measurement application.

  • Stability for around 10,000 loading/unloading cycles.6

  • Response time is less than 11 ms.5

  • Operating voltage is 0.01 V.7

48.0 MXenes
Nanosheets (Ti3C2)
Piezoresistive - 0.351 7.5 [250]
  • No blood pressure measurement application.

  • Stability for around 4000 loading/unloading cycles. 6

  • Response time is < 30 ms. 5

  • Operating voltage is 1.0 V. 7

49.0 Pt-coated
Polymeric Nanofibers
(Nanohair)
Piezoresistive 0.005 - 11.35 [293]
  • No blood pressure measurement application.

  • Response time is 50 ms. 5

  • Stability for around 10,000 loading/unloading cycles. 6

50.0 ITO (3D Printed Mold) Piezocapacitive 0.003 4.0 1.62 [294]
  • No blood pressure measurement application but heart pulse.

  • Stability for around 1000 loading/unloading cycles. 6

  • Operating voltage can be 3.0 V. 7

  • Response time is < 10 ms. 5

51.0 Ag Flexible Piezoelectret-Based Pressure Sensor Piezoelectric - 2.5 15.0 [295]
  • No blood pressure measurement application.

  • Stability for around 36,000 loading/unloading cycles with constant pressure. 6

  • Perfluoro (alkoxy alkane) electret.

  • Not all properties were reported.

52.0 ITO/PDMS
(Pattered micro-pyramid)
Piezocapacitive 0.003 20.0 0.55 [296]
  • No blood pressure measurement application.

  • Response time is 1.0 s. 5

  • Stability for around 10,000 loading/unloading cycles. 6

  • The gate voltage is higher than 5.0 V. 7

53.0 Ag NWs (Embedded PDMS Electrode with Microarray Structure) Piezocapacitive 0.003 5.0 2.94 [297]
  • No blood pressure measurement application.

  • Stability for around 1000 loading/unloading cycles. 6

  • Response time is less than 50 ms. 5

  • Not all properties were reported.

1 The smallest change that can be detected by a sensor. 2 The largest change that can be detected by a sensor. 3 The rate of change in output signal when responding to a change in stimulus. 4 Calculated Sensitivity. 5 The time interval within which a sensor can detect the smallest change in stimulus. 6 The rate of change in sensitivity and/or other sensor properties over continuous cycles of loading/unloading. 7 The sensor operating voltage should be as small as fractions of an mV to reduce power consumption, improve the battery life of the wearable system and/or support the integration with other energy harvesting applications.