Table 4. Performance of the Firefly technology as a screener for a clinical threshold of AHI ≥15.
Performance (AHI ≥15) | Training set | Test set |
---|---|---|
Sensitivity (%) (CI) | 89.7 (75.4–97.3) | 88.3 (77.3–95.1) |
Specificity (%) (CI) | 86.5 (77.9–92.3) | 80.0 (67.8–88.7) |
PPV (%) (CI) | 74.5 (59.5–85.7) | 81.5 (70.9–89.6) |
NPV (%) (CI) | 95.1 (88.5–98.7) | 87.3 (76.1–94.4) |
Accuracy (%) (CI) | 87.5 (81.3–92.3) | 84.2 (75.8–89.2) |
ROC AUC (CI) | 0.95 (0.91–0.98) | 0.92 (0.85–0.95) |
PR AUC (CI) | 0.87 (0.78–0.94) | 0.89 (0.80–0.94) |
Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), accuracy and AUC of the ROC and PR curves are provided for the training and testing sets respectively. The 95% confidence intervals are shown in parenthesis. AHI, apnea hypopnea index; ROC, receiver operating characteristic; AUC, area under the curve; PR, precision recall.