TABLE II.

Performance Comparison of the Proposed Pulse Similarity Index-Based Algorithm and a Previous Activity Index-Based Algorithm [14] for Artifact Rejection at Different Levels and Types of Artifacts

Algorithms’ Performance (%) Proposed Pulse Similarity Index-based Algorithm Activity Index-based Algorithm [14]
Overall Accuracy 96.25 86.50
Overall Error 03.75 13.50
Sensitivity 92.51 73.35
Specificity 97.35 90.36
Positive predictive value 91.11 69.09
Negative predictive value 97.79 92.03
Accuracy (Below Normal Amplitude) 95.74 90.63
Accuracy (Normal Amplitude) 95.10 74.51
Accuracy (Above Normal Amplitude) 100 100
Accuracy (Low-Level Movement) 93.83 72.84
Accuracy (High-Level Movement) 96.81 94.15
Accuracy (Electrode Detachment) 100 100