Table 1.
parameter | clinical units | S.I. Units | references |
---|---|---|---|
ventilator parameters | |||
PEEP | 5 cmH2O | 490 Pa | [17,18] |
PIP (reference value) | 15 cmH2O | 1471 Pa | [17,18] |
RR | 15 breaths/min | [17,18] | |
I : E ratio | 1 : 2 | [17,18] | |
splitter parameters | |||
RI | 0.06 cmH2O/L/s | 5670 Pa s m−3a | derived |
RV | variable | variable | N/A |
RE | 0.06 cmH2O/L/s | 5670 Pa s m−3a | derived |
patient parameters | |||
RL | 2 cmH2O/L/s | 196133 Pa s m−3 | [19] |
CL (reference value) | 0.054 L/cmH2O | 5.5 × 10−7 m3 Pa−1 | [17] |
RETT | 8.0 cmH2O/L/s | 784000 Pa s m−3 | [19] |
aResistance (R) derived from Poiseuille law for straight laminar pipe, R = (128 µL)/(πD4), assuming diameter D = 22 mm, viscosity of air µ = 18.13 × 10−6 Pa s, length L = 1.8 m. PEEP, positive end-expiratory pressure; PIP, positive inspiratory pressure; RR, respiratory rate; I : E ratio, inspiratory/expiratory ratio. Reference values represent healthy patients and are used for Lung Model A, defined below.