. 2020 Aug 24;7(8):200585. doi: 10.1098/rsos.200585

Table 3.

Achieved tidal volumes during pressure control ventilation for paired patients with mismatched lung compliance using the standard splitter and PIP was adjusted to achieve a tidal volume of approximately 490 ml in Patient 1.

Patient 1–Patient 2 PIP (cmH2O) Patient 1 tidal volume (ml) Patient 2 tidal volume (ml)
A–A 15 490.2 490.2
A–B 15 490.3 410.9
A–C 15 490.3 366.2
A–D 15 490.3 318.4
B–C 17 493.2 449.5
B–D 17 493.2 382.1
C–D 18.5 494.5 429.9