Skip to main content
. 2020 Aug 24;7(8):200585. doi: 10.1098/rsos.200585

Table 4.

Achieved tidal volumes during pressure control ventilation for paired patients with mismatched lung compliance using the modified splitter and with patient-specific PIP and variable resistance settings.

Patient 1–Patient 2 PIP (cmH2O) Patient 1 variable resistor (cmH2O/L/s) achieved tidal volume
Patient 1 Patient 2
A–A 15 0 490.2 490.2
A–B 17 7 496.4 492.9
A–C 18.33 11.67 492.9 488.1
A–D 20.33 18 492.5 488.0
B–C 18.5 6.67 492.8 494.2
B–D 20.3 11.94 494.0 494.2
C–D 20.3 8.67 494.3 488.1