Skip to main content
. 2020 Aug 24;7(8):200585. doi: 10.1098/rsos.200585

Table 5.

Values of Patient 1 and Patient 2 inhalation pathway variable resistance values, ventilator PIP and resulting tidal volume for Patient 1 and Patient 2 when adjusting variable resistors and PIP to achieve equivalent tidal volume (C–C), lower tidal volume in Patient 2 (C–C(−)) and higher tidal volume in Patient 2 (C–C(+)). Tidal volume in Patient 1 is maintained throughout.

Patient 1–Patient 2 P1 variable resistor (cmH2O/L/s) P2 variable resistor (cmH2O/L/s) PIP (cmH2O) tidal volume (ml)
Patient 1 Patient 2
C–C 0 0 18.5 494.5 494.5
C–C(−) 0 16.67 18.5 494.2 379.3
C–C(+) 16.67 0 22.67 496.5 646.9