Table 1.

Categories of failure for IROC lung and SBRT spine phantom irradiations from Jan 2012 – Dec 2018, along with head and neck phantom failure data from a previous IROC study[5] (Nov 2014 – Oct 2015).

Error Type Number of phantoms (%) 95% C.I. (%)
Lung

Systematic dose 28 (22%) (17,33)
  ➣ overdose 9 (6%)
  ➣ underdose 19 (16%)
Local dose 7 (6%) (3, 12)
Localization: major motion 64 (50%) (46,64)
  ➣ superior 26 (20%)
  ➣ inferior 38 (30%)
Localization: minor motion (ant-post) 8 (6%) (4,13)
Localization: no motion (left-right) 7 (6%) (3,12)
Global 3 (2%) (1,7)
Combination category 10 (8%) (5,15)

Spine

Systematic dose 25 (60%) (44,73)
  ➣ overdose 9 (22%)
  ➣ underdose 16 (38%)
Dose fall-off region 5 (12%) (5,25)
OAR overdose 6 (14%) (7,28)
Localization 6 (14%) (7,28)

Head and Neck [5]

Systematic dose 32 (62%) (48,74)
  ➣ overdose 6 (12%)
  ➣ underdose 26 (50%)
Local dose 8 (15%) (8,28)
Localization 6 (12%) (5,23)
Global 6 (12%) (5,23)