TABLE 1.

Statistical analysis of algorithm determined cardiac output (ih + ip) and systemic vasculature resistance (R) compared against the thermodilution measured cardiac output (TCO) and systemic vasculature resistance (SVR).

Cardiac output (L/min) Systemic vascular resistance (dynes-s/cm5)
Reference mean (low-high) 6.0 (3–10.6) 1054 (501–1897)
Bias (mean) − 0.11 − 62
Precision (standard deviation) 0.89 161
Limits of agreement (95% confidence interval) 1.75 315
Percentage error 29.1% 29.9%
Correlation plot (slope) 1.08 1.04
Correlation plot (offset) − 0.39 18.8
Correlation coefficient (r) 0.82 0.86

Correlation statistics for cardiac output measurements calculated from data presented in Fig. 4. Correlation statistics calculated for systemic vascular resistance measurements calculated from data presented in Fig. 5.