Table 2.

Intra-and inter-day precision and accuracy for QC samples of silmitasertib in various matrices (Intra-day: n=6; Inter-day: n=18)

Matrix Nominal
a(ng/ml)
Mean
conc.
a(ng/ml)
Within-day
(n= 6)
Mean
conc.
a(ng/ml)
Between-day
(n= 18)
%CV %Nominal %CV %Nominal
Human plasma (0.2-125 ng/ml) 0.2 0.199 4.3 99.4 0.204 6.3 101.9
0.6 0.603 5.4 100.5 0.600 4.6 99.9
7.5 7.57 2.9 100.9 7.32 3.3 97.6
100 94.6 2.5 94.6 93.6 2.1 93.6
Human plasma (32-20,000 ng/ml) 32 31.66 2.4 98.9 32.76 8.2 102.4
96 91.93 2.8 95.8 95.28 4.1 99.2
1300 1260.4 2.5 97.0 1318.8 5.1 101.4
17,000 15,168.6 1.9 89.2 15,482.8 3.0 91.1
Human CSF 0.2 0.213 1.9 106.3 0.210 5.2 107.0
0.6 0.590 4.4 98.3 0.610 5.7 101.7
3 2.890 3.5 96.5 2.93 4.8 97.5
18 17.55 4.7 97.5 17.34 4.5 96.3
Mouse brain 2 1.73 5.0 86.4 1.82 10.7 91.0
4 3.58 5.3 89.4 3.84 7.2 96.1
16 15.01 3.1 93.8 15.92 6.3 99.5
32 28.70 1.7 89.7 31.34 9.2 97.9
bPrimate brain 4 4.32 3.2 108.0 4.29 4.4 107.2
32 31.59 2.3 98.7 32.42 5.1 101.3
a

The unit of concentration is ng/g for QC samples in mouse brain and primate brain

b

QC samples for primate brain were analyzed by using the calibration curve of mouse brain homogenate.