Skip to main content
. 2020 May 22;21(8):278–288. doi: 10.1002/acm2.12906

Table 4.

Face‐on angular dependence of microDiamond for 6X square fields upon a Varian Clinac IX. Angles include range from 240° to 60° normalized to 0° measurement

240° to 60° 0.5 × 0.5 cm2 1 × 1 cm2 2 × 2 cm` 3 × 3 cm2 10 × 10 cm2
Maximum response 1.21 1.16 1.15 1.14 1.10
Minimum response 0.99 0.98 0.98 0.98 0.98
Range 0.22 0.18 0.17 0.16 0.12