Skip to main content
. 2020 May 22;21(8):278–288. doi: 10.1002/acm2.12906

Table 5.

Edge‐on angular dependence of microDiamond for ± 120° around 0°. The PTW microDiamond is continuously in edge‐on orientation for all gantry angles for 6X square field upon a Varian Clinac IX

0° ± 120° 0.6 × 0.6 cm2 1 × 1 cm2 2 × 2 cm2 3 × 3 cm2
Maximum response 1.01 1.03 1.01 1.00
Minimum response 0.72 0.94 0.99 0.98
Range 0.28 0.08 0.02 0.02