Skip to main content
. 2020 May 22;21(8):278–288. doi: 10.1002/acm2.12906

Table 6.

Percentage increase in output factor for a range of rectangular fields positioned along the cable of the microDiamond relative to when positioned perpendicular to the cable

Field size (cm2) Stem effect (%)
6X 6FFF
1 × 10 0.83 0.93
3 × 10 0.19 0.17