. 2020 May 22;21(8):278–288. doi: 10.1002/acm2.12906

Table 7.

TPR2020,10 (S) measured with the CC01, microDiamond in edge‐on orientation and Razor Diode for a 6X beam on a Clinac XI for a range of square field sizes. The percentage difference between the microDiamond to CC01 and Razor diode to CC01 are also included

Square field size cmn CC01 microDiamond % diff to CC01 Razor diode % diff to CC01
0.5 × 0.5 0.585 0.590 0.80 0.600 2.61
1 × 1 0.613 0.613 0.00 0.615 0.33
1.5 × 1.5 0.618 0.620 0.32 0.621 0.49
2 × 2 0.620 0.623 0.48 0.623 0.48
3 × 3 0.626 0.625 −0.16 0.627 0.16
4 × 4 0.629 0.629 0.00 0.632 0.48
5 × 5 0.633 0.635 0.32 0.637 0.63
6 × 6 0.639 0.640 0.16 0.643 0.63
8 × 8 0.650 0.650 0.00 0.653 0.46
10 × 10 0.661 0.660 −0.15 0.664 0.45