Figure 2.

Figure 2

Plot of data collected from 10 volunteers during a “breathe down test” (see Methods). Each point on the plot represents one pair of a phone reading and Masimo Radical 7 reading. The root-mean-square deviation of bias between the two instruments in this test was 2.2 oxygen saturation counts. The upper and lower limits of agreement per Bland and Altman20 are shown as dashed lines. A second-order polynomial fit of the data is shown as a solid line. Spo2 = oxygen saturation.