Table 1.

Details of the contrast medium specifics

Phantom size Scan parameter tube voltage: Constant CTDI Dose modulation
70 kVp 80 kVp 100 kVp 120 kVp 140 kVp 70 kVp 80 kVp 100 kV 120 kV 140 kV
Large Rotation time (s) 1 1 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5
mAs (range 80–500) 500 395 405 250 170 499 499 464 372 282
CTDIvol (mGy) 13.17 16.6 16.68 16.6 16.43 6.59 10.11 14.79 16.6 17.59
Medium Rotation time (s) 1 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5
mAs (range 80–500) 380 475 240 150 105 499 499 322 244 178
CTDIvol (mGy) 10 9.98 9.89 9.96 10.15 6.09 8.14 8.98 10 11.6
Small Rotation time (s) 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5
mAs (range 80–500) 500 330 170 105 70 484 630 219 106 123
CTDIvol (mGy) 6.59 6.93 7.00 6.97 6.77 4.67 5.18 5.73 7.05 8.75

CTDIvol Volume computed tomography dose index. Scan technique: constant CTDIvol and dose modulation (SmartmA) using conventional, iterative reconstruction. Additional parameters: scan mode, helical; detector collimation, 80 × 0.625; pitch, 0.5; scan field of view, 50 cm; display field of view, 36 cm; reconstructed slice thickness, 2.5 mm; reconstruction kernel, standard; iterative reconstruction, adaptive statistical iterative reconstruction, ASIR-V 50%; noise index14.5 (scans with dose modulation)