Skip to main content
. 2020 Aug;10(4):965–981. doi: 10.21037/cdt.2020.01.08

Table 1. A typical carotid CEUS protocol.

Parameter Reference value/setting
MI <0.1
UCA dose 2.4 mL of SonoVue®, followed by 5–10 mL of normal saline
Injection pattern Bolus injection usually performed
Infusion technique with special injector is also available for prolonged time of enhancement
Injection site Usually antecubital fossa vein
Central venous lines and ports can also be used
Cannula required A 20-gauge cannula can be safely used to avoid microbubbles disruption (18–21-gauge cannulas can result in similar enhancement) (15)
CEUS technique Pulse-inversion technique should be used although other techniques are also available in different manufacturers (e.g., amplitude modulation)
Appearance of microbubbles This depends on vessel examined and the patient’s circulatory status but microbubbles usually appear in the carotid arteries in 20–30 seconds
Duration of enhancement Depending on scanning parameters, the enhancement may last up to 4 minutes
General comment Albeit of UCA excellent safety profile, resuscitation equipment and trained personnel should always be available for treatment of allergic reactions

CEUS, contrast-enhanced ultrasound; MI, mechanical index; UCA, ultrasonographic contrast agent.