Comparison of histopathology measurements of the resected specimens with MRI measurements using three-dimensional (3D) and axial measurements according to RECIST

Sequence (n = cases)a Median (IQR) histopathological diameter (cm) Median (IQR) MRI diameter (cm) Reader 1 Median (IQR) difference (cm) Reader 1 Median (IQR) Absolute error (cm) Reader 1 p-value Reader 1
Reader 2 Reader 2 Reader 2 Reader 2
Axial T1w precontrast (n = 31) 5.8 (4.0) 4.7 (3.3) −1.4 (1.8) 1.1 (1.7) < 0.001
4.5 (3.6) −0.8 (1.8) 1.1 (1.7) < 0.001
3D T1w precontrast (n = 31) 5.8 (4.0) 5.4 (5.1) −0.3 (1.3) 0.7 (0.7) 0.262
5.3 (5.2) −0.3 (1.3) 0.8 (0.9) 0.159
Axial T1w postcontrast (n = 37) 5.8 (5.2) 4.9 (4.3) −1.0 (1.9) 1.1 (1.5) < 0.001
5.0 (4.4) −0.9 (1.3) 1.0 (1.2) < 0.001
3D T1w postcontrast (n = 37) 5.8 (4.2) 5.5 (4.8) −0.1 (1.2) 0.6 (0.6) 0.739
5.5 (4.9) −0.2 (1.4) 0.7 (0.9) 0.722
Axial T2w TSE (n = 38) 5.8 (4.2) 4.7 (4.2) −1.0 (1.6) 1.3 (1.7) < 0.001
4.8 (4.2) −1.0 (1.6) 1.2 (1.6) < 0.001
3D T2w TSE (n = 38) 5.8 (4.2) 5.4 (5.1) −0.3 (1.1) 0.6 (0.8) 0.379
5.4 (5.1) −0.1 (1.1) 0.5 (0.7) 0.722
Axial DWI (n = 26) 6.5 (4.9) 5.5 (6.0) −1.3 (2.2) 1.4 (2.0) < 0.001
5.5 (6.0) −1.2 (2.2) 1.4 (2.0) < 0.001
3D DWI (n = 26) 6.5 (4.9) 6.5 (5.4) −0.2 (1.3) 0.7 (1,2) 0.159
6.4 (5.3) −0.2 (1.5) 0.8 (1.1) 0.196

aNot all sequences in all patients acquired