Table 1.
Interpretation of Results for Patient Samples Based on Data from Either Singleplex Assay
N1 | N2 | RP | Results Interpretation | Report | Follow-Up Action |
---|---|---|---|---|---|
+ | + | −/+ | SARS-CoV-2 detected | POSITIVE | report into electronic medical record (EMR) |
If only one of these targets is detected | −/+ | SARS-CoV-2 detected | POSITIVE | report into EMR | |
If signal is detected, but no target reaches 2/2 detected technical repeats | + | inconclusive | INCONCLUSIVE | sample is repeated at qRT-PCR step once more. If sample is still inconclusive, the result is reported into the EMR as such and it is recommended that a new sample is obtained from the patient | |
− | − | + | SARS-CoV-2 not detected | NEGATIVE | report into EMR |
− | − | − | Invalid | INVALID | the result is reported into the EMR as such and it is recommended that a new sample is obtained from the patient |