Interpretation and Purpose of Controls for Two- and One-Step Singleplex Assays

Control Type Control Name Used to Monitor/Where to Repeat Protocol Expected Results
Expected CT Values
N1 N2 RP
Negative NTC contamination during qRT-PCR process undetected
Positive SARS-CoV-2 RNA amplification/primer-probe integrity + + <40 (N1, N2,) and undetected (RP)
Extraction POS EC extraction/sample stability + + + <40 (N1, N2, RP)
NEG EC extraction cross-contamination + undetected (N1, N2) and <40 (RP)