Table 4.
Interpretation and Purpose of Controls for the One-Step Multiplex Assay
| Control Type | Control Name | Used to Monitor/Where to Repeat Protocol | Expected Results |
Expected CT Values | |||
|---|---|---|---|---|---|---|---|
| N | S | ORF1ab | MS2 | ||||
| Negative | NTC | contamination during qRT-PCR process | - | - | - | - | Undetected |
| Positive | TaqPath Combo Kit positive control | amplification/primer-probe integrity | + | + | + | - | <40 (N, S, ORF1ab) and undetected (MS2) |
| Extraction | POS EC | extraction/sample stability | + | + | + | + | <40 (N, S, ORF1ab, MS2) |
| NEG EC | extraction cross-contamination | - | - | - | + | undetected (N, S, ORF1ab) and <40 (MS2) | |
| phage MS2 gene | extraction/amplification/sample stability | NA | NA | NA | + | <40 (MS2) | |