Table 4.

Interpretation and Purpose of Controls for the One-Step Multiplex Assay

Control Type Control Name Used to Monitor/Where to Repeat Protocol Expected Results
Expected CT Values
N S ORF1ab MS2
Negative NTC contamination during qRT-PCR process - - - - Undetected
Positive TaqPath Combo Kit positive control amplification/primer-probe integrity + + + - <40 (N, S, ORF1ab) and undetected (MS2)
Extraction POS EC extraction/sample stability + + + + <40 (N, S, ORF1ab, MS2)
NEG EC extraction cross-contamination - - - + undetected (N, S, ORF1ab) and <40 (MS2)
phage MS2 gene extraction/amplification/sample stability NA NA NA + <40 (MS2)