Typical parameters for 2D-DARR experiments.

MAS rate: 10 kHz (rotor period, τr = 100 μs)
Pulse Power
(kHz)
Duration
(μs)
Phase
pH,1 110 2.27 (x)8, (−x)8
pH,2 Linear ramp between 40 and 80 2,000 y
pC,1 50 2,000 y
pC,2 (π/2) 50 5 x,−x and y,−y for each t1
pC,3 50 5 x,x,y,y,−x,−x,−y,−y
ϕReceiver x,−x,y,−y,−x,x,−y,y,−x,x,−y,y,x,−x,y,−y
Delay Duration (s)
d1 2
t1 varied for 2D acquisition
Mixing time Duration (ms)
τ 10-1500
Decoupling period Power (kHz) Method
dec1 110 TPPM
Comments During mixing, 1H continuous wave radio frequency should be applied with power equal to the MAS rotation rate (10 kHz).