Table 7:
Typical parameters for 2D ZF-TEDOR experiments.
MAS rate: 9 kHz (rotor period, τr =~110 μs) | |||
---|---|---|---|
Pulse | Power (kHz) | Duration (μs) |
Phase |
pH,1 (π/2) | 110 | 2.27 | (x)16, (−x)16 |
pH,2 | Linear ramp: 39.33 to 78.67 | 2,000 | x |
pC,1 | 50 | 2,000 | x |
pC,2 (π) | 50 | 10 | x,x,y,y |
pC,3 (π/2) | 50 | 5 | x |
pC,4 (π/2) | 50 | 5 | y,y,−y,−y |
pC,5 (π/2) | 50 | 5 | (x)4,(y)4,(−x)4,(−y)4 |
pN,1 (recoupling) | 50 | 10 | −y,y,y,−y,x,−x,−x,x,y,−y,−y,y,−x,x,x,−x,y,−y,−y,y,−x,x,x,−x,−y,y,y,−y,x,−x,−x,x |
pN,2 (π/2) | 50 | 5 | x,−x |
pN,3 (π/2) | 50 | 5 | x,−x |
ϕReceiver | −y,y,y,−y,x,−x,−x,x,y,−y,−y,y,−x,x,x,−x,y,−y,−y,y,−x,x,x,−x,−y,y,y,−y,x,−x,−x,x | ||
Delay | Duration (s) | ||
d1 | 2 | ||
t1 | varied for 2D acquisition | ||
τzf | 0.0002 | ||
Decoupling period | Power (kHz) | Method | |
dec1 | 110 | TPPM | |
dec2 | 110 | TPPM | |
dec3 | 110 | CW | |
comments | Two pN,1 pulses are applied per rotor period. The number of pN,1 applied at a time depends on the choice of tmix, which is typically between 1 and 20 ms. |