Table 4.

Evaluation of the accuracy and precision of the plate count and TEMPO® methods at concentrations close to the acceptance criteria for human milk before pasteurization.

Expected Measured CV (%) Accuracy¥
concentration concentration (%)
(log CFU/ml) (log CFU/ml)
Plate count method
Total aerobic flora 5.3 5.2 ± 0.1 1.7 98.1
Enterobacteriaceae 4.0 4.0 ± 0.1 1.4 100.0
S. aureus 3.0 2.7 ± 0.1 2.0 90.0
Bacillus cereus (spores) 1.0 0.8 ± 0.6 81.6 80.0
TEMPO® method
Total aerobic flora 5.3 5.3 ± 0.1 2.9 100.0
Enterobacteriaceae 4.0 4.1 ± 0.1 3.0 102.5
S. aureus 3.0 2.5 ± 0.1 5.3 83.3
Bacillus cereus (spores) 1.0 1.2 ± 0.1 9.1 120.0
¥

Accuracy = (measured concentration/expected concentration) ×100 (n = 3).

Coefficient of variation = (standard deviation/mean) ×100 (n = 6).