Table 3.

Overall performance characteristics of Care Start™ in comparison with the standard light microscopy, Arba Minch, 2016.

Performance characteristics Care Start™ malaria
Sensitivity (95% CI) 97.44% (91.04%–99.69%)
Specificity (95% CI) 93.67% (85.84%–97.91%)
Positive predictive value (95% CI) 93.83% (86.18%–97.97%)
Negative predictive value (95% CI) 97.37% (90.82%–99.68%)