Skip to main content
. 2020 Aug 28;5(36):23351–23357. doi: 10.1021/acsomega.0c03229

Table 1. Validation Results for CN and ATCA by PESI/MS/MS (n = 5).

          low QC
high QC
intraday
interday
intraday
interday
compounds range (ng/mL) R2 LLOD (ng/mL) LLOQ (ng/mL) precision (%) accuracy (%) precision (%) accuracy (%) precision (%) accuracy (%) precision (%) accuracy (%)
CN 50–2500 0.997 42 127 250 ng/mL 2000 ng/mL
7.8 1.9 5.6 1.1 5.1 5.2 15 14
ATCA 50–1500 0.995 43 131 250 ng/mL 1000 ng/mL
8.3 4.8 7.0 5.9 9.7 3.3 5.4 3.8