Skip to main content
. 2020 Mar 7;52(2):407–417. doi: 10.1002/jmri.27106

Table 1.

Scan Parameters for qT2, DWI, Dixon, and Fat‐Suppressed T2‐Weighted Sequences

qT2 DWI Noise map Dixon T2w‐SPIR
Sequence MSE SE‐EPI SE‐EPI MS_FFE MS_TSE
Matrix size 160 × 160 160 × 160 160 × 160 320 × 320 640 × 640
Voxel size (mm3) 3 × 3 × 6 3 × 3 × 6 3 × 3 × 6 1.5 × 1.5 × 6 0.75 × 0.75 × 3
Slice gap (mm) 6
Slices 17 31 31 31 31
TR (msec) 5633 5000 210 5500
TE (msec) 17*7.6 57 57 2.6/3.36/4.12/4.88 70
Flip angle 90/180° 90/180° 90/180° 90°
b‐value (s/mm2) (# gradient orientations) 0(1), 1(6), 10(3), 25(3), 100(3), 200(6), 400(8), 600(12) 0(1), 400(1)
Fat suppression SPAIR/SPIR/SSGR SPAIR/SPIR/SSGR SPIR
SENSE 2 1.9 1.9 2 2

SENSE: sensitivity encoding.