Skip to main content
. 2020 Aug 13;412(26):7131–7140. doi: 10.1007/s00216-020-02846-8

Table 1.

Comparison of sensitivities and LODs for DC-mode and RF-mode of all investigated VOCs. LODs were calculated for 1 s of integration time. The DC-mode data were collected at Edrift = 66 V/cm, RF = 0 Vp-p, and DC = 66 V/cm under humid conditions. The RF-mode data were collected at Edrift = 66 V/cm, RF = 120 Vp-p, and DC = 13.5 V/cm under humid conditions

Sensitivity (cps ppbV−1) LOD (ppbV)
Compound (m/z) DC-mode RF-mode RF-mode/DC-mode DC-mode RF-mode RF-mode/DC-mode
Methanol (33) 27.5 116.5 4.2 0.735 0.432 1.7
Acetonitrile (42) 46.7 418.2 8.6 0.337 0.11 3.1
Acetaldehyde (45) 56.5 458.1 8.1 0.66 0.245 2.7
Ethanol (47) 3.4 18.2 5.3 12.319 8.31 1.5
2-Propenal (57) 43 406.3 9.4 0.301 0.146 2.1
Acetone (59) 103.6 1008.3 9.7 0.26 0.09 2.9
Isoprene (69) 10 120.8 12.1 1.226 0.463 2.6
2-Butenal (71) 65.8 701 10.7 0.163 0.069 2.4
2-Butanone (73) 66.1 750.9 11.4 0.325 0.104 3.1
Benzene (79) 39.2 423.6 10.8 0.22 0.056 3.9
Toluene (93) 50.2 620.3 12.4 0.144 0.06 2.4
o-Xylene (107) 65.4 747.4 11.4 0.115 0.03 3.8
Chlorobenzene (113) 38.2 413.7 10.8 0.155 0.061 2.5
α-Pinene (137) 25.7 252.1 9.8 0.265 0.087 3
1,2-Dichlorobenzene (147) 43.9 264.7 6.0 0.087 0.044 2