Skip to main content
. 2020 Jun 22;65(7):3373–3383. doi: 10.1021/acs.jced.0c00184

Figure 6.

Figure 6

(Top) Calibration results for signal ratio of oxygen to nitrogen vs set mole fraction of ethanol at different pressures and temperatures (errorbars: standard-deviation of 32 single spectra); (bottom) deconvolution of range = (1330 to 1570) cm–1 for calibration spectra of different mole fractions at T = 522 K and p = 8 MPa.