Gamma Passing rates for the measured data, data that has been shifted in the LR direction, and data that has been scaled by 3%.

Gamma passing rate (3%/1 mm) Gamma passing rate (1%/1 mm)
Original 1 mm lateral Shift 3% positive dose scaling Original 1 mm lateral Shift 3% positive dose scaling
99.4 96.2 98.6 65.1 62.1 44.9
100 97.1 94.4 97.7 87.4 74.6
99.4 96.6 99.3 91.9 83.7 84.2
99.8 98.6 98.2 88.1 77.8 64
99.9 98.9 98.9 94.9 83.7 80.6
99.7 98.3 99.3 86.7 81.7 67.6
98.9 97.7 97.2 63 59.9 38.7
99.4 96.8 98.2 69.3 60.7 46.3
99.5 98 99.4 95.6 90.6 91.6
98 95 97.3 87.5 78.8 79.4
98.2 94.5 90.1 88.3 81.2 60.5
98.1 92.6 97.8 83.9 67.7 80.7
100 98.9 99.9 100 91.9 99.2
99.5 95.1 99.6 96.9 82.7 98.2
99.7 97.3 97.8 91.4 81 77
Average 99.3 96.8 97.7 86.7 78.1 72.5
Standard deviation 0.68 1.82 2.51 11.78 10.51 18.73