Skip to main content
. 2020 Jul 28;21(9):201–208. doi: 10.1002/acm2.12974

Table 2.

Measured contrast per fructose concentration, tube current, and reconstruction method. Mean ± SD values of 45 measurements are shown (five repeated acquisitions per condition × 9 images per acquisition).

Fructose concentration (mg/mL) FBP AIDR 3D
10 mA 20 mA 30 mA 40 mA 100 mA 120 mA 10 mA 20 mA 30 mA 40 mA 100 mA 120 mA
240 38.9 ± 4.5 37.7 ± 2.2 38.1 ± 1.9 37.0 ± 1.5 36.5 ± 0.9 36.4 ± 0.8 39.3 ± 2.2 37.6 ± 1.2 37.5 ± 1.2 37.0 ± 1.0 36.9 ± 0.8 36.7 ± 0.7
250 35.9 ± 4.7 34.8 ± 3.0 34.7 ± 2.2 34.8 ± 1.4 33.9 ± 0.9 33.5 ± 1.0 36.0 ± 2.1 35.1 ± 1.6 34.9 ± 1.6 34.7 ± 1.0 34.2 ± 0.8 33.8 ± 0.9
260 28.8 ± 4.2 30.8 ± 2.9 30.7 ± 1.9 29.6 ± 2.3 29.0 ± 1.0 29.1 ± 0.9 30.3 ± 2.6 30.7 ± 1.7 30.5 ± 1.4 29.9 ± 1.8 29.4 ± 0.9 29.5 ± 0.9
280 23.8 ± 5.2 24.1 ± 2.9 23.7 ± 2.5 23.6 ± 1.5 23.1 ± 1.2 22.7 ± 1.3 24.8 ± 2.4 24.6 ± 1.7 23.6 ± 2.0 23.9 ± 1.1 23.5 ± 1.1 23.2 ± 1.1
290 19.6 ± 4.1 21.2 ± 2.5 20.0 ± 2.2 19.7 ± 1.4 19.4 ± 1.0 19.3 ± 1.2 21.0 ± 1.6 21.5 ± 1.4 20.2 ± 1.6 20.0 ± 1.2 19.8 ± 0.9 19.7 ± 1.1
300 18.5 ± 5.9 18.7 ± 2.2 18.3 ± 2.3 16.8 ± 1.8 17.1 ± 1.3 17.3 ± 1.0 19.6 ± 2.5 18.8 ± 1.1 18.6 ± 1.6 17.5 ± 1.6 17.6 ± 1.2 17.8 ± 1.0
310 13.2 ± 5.5 13.7 ± 2.6 13.4 ± 2.4 12.6 ± 1.5 12.8 ± 1.2 12.7 ± 1.0 14.3 ± 1.7 14.2 ± 1.5 14.0 ± 1.4 13.2 ± 1.2 13.3 ± 1.2 13.2 ± 0.9
320 10.8 ± 4.5 12.1 ± 3.2 11.0 ± 2.1 10.8 ± 1.4 10.8 ± 1.1 10.3 ± 1.1 12.3 ± 2.2 12.4 ± 1.7 11.7 ± 1.5 11.5 ± 1.0 11.3 ± 1.0 10.8 ± 1.1
330 9.8 ± 4.3 9.0 ± 2.0 8.5 ± 2.4 8.1 ± 1.8 7.8 ± 1.6 8.0 ± 1.2 10.7 ± 1.6 9.8 ± 1.5 9.3 ± 1.7 8.8 ± 1.5 8.3 ± 1.5 8.6 ± 1.1
340 2.6 ± 3.6 4.4 ± 2.7 3.9 ± 2.2 3.4 ± 1.7 3.8 ± 1.2 3.8 ± 1.2 5.0 ± 1.7 5.7 ± 1.4 4.8 ± 1.7 4.4 ± 1.5 4.3 ± 1.1 4.3 ± 1.2