Comparison of gel‐measured and calculated dose–volume metrics, D95 for different treatment scenarios.

Target Estimated D95 or D100 values
TPS Reference Scenario 1 Scenario 2
Measured Difference Measured Difference Measured Difference
PTV1 (D95) 62.0% 67.6% 5.6% 66.6% 4.7% 65.1% 3.1%
PTV2 (D95) 76.3% 78.1% 1.9% 76.2% 0.0% 77.5% 1.3%
GTV2 (D100) 70.5% 73.1% 2.6% 70.5% 0.0% 72.1% 1.6%