. 2020 Aug 25;21(9):244–251. doi: 10.1002/acm2.13012

Table 1.

Variability of detector response when changing its orientation with respect to the magnetic field (B) force lines.

Detector orientation with respect to B
0 90 270
Measurement 1 (counts) 8,11E+06 8,05E+06 8,04E+06
Measurement 2 (counts) 8,11E+06 8,05E+06 8,06E+06
Measurement 3 (counts) 8,13E+06 8,07E+06 8,09E+06
Mean value (counts) 8,12E+06 8,06E+06 8,08E+06
Standard deviation 1,26E+03 1,37E+03 3,04E+03
Coefficient of variation 0,16% 0,29% 0,37%
Mean dose (Gy) 1,00 0,993 0,995