Table 5.
Precision analysis of the SARS-COV-2 antigen quantitative detection kit (ELISA).
Within-day precision | Day to day precision | |||||
---|---|---|---|---|---|---|
S1 | S2 | S3 | S1 | S2 | S3 | |
Mean value(pg/mL) | 6.49 | 41.15 | 163.10 | 6.51 | 41.13 | 166.33 |
SD(pg/mL) | 0.56 | 2.68 | 9.46 | 0.62 | 3.07 | 10.36 |
CV(%) | 8.65% | 6.51% | 5.80% | 9.52% | 7.47% | 6.23% |