Table 3.

Intra-session test-retest reliability of the DynaPort MT (n = 40).

Parameter Pretest
(Mean ± SD)
Retest
(Mean ± SD)
Bias
(95% CI)
95% LOA ICC2,1
(95% CI)
SEM
(SEM%)
MDC95
(MDC95%)
Gait speed [m/s] 0.60 ± 0.18 0.60 ± 0.20 0.00
(–0.03 to 0.03)
–0.19 to 0.18 0.89
(0.79 to 0.94)
0.06 (10.6) 0.18 (29.3)
Cadence [steps/min] 90.1 ± 15.3 91.7 ± 16.7 –1.7
(–3.7 to 0.3)
–13.9 to 10.6 0.92
(0.86 to 0.96)
4.4 (4.9) 12.2 (13.5)
Step length [cm] 39.4 ± 9.4 38.9 ± 9.7 0.5
(–1.2 to 2.3)
–10.1 to 11.1 0.84
(0.72 to 0.91)
3.8 (9.8) 10.6 (27.0)
Step time [s] 0.69 ± 0.13 0.68 ± 0.13 0.01
(–0.01 to 0.03)
–0.12 to 0.10 0.91
(0.84 to 0.95)
0.04 (6.5) 0.12 (18.0)
Step time asymmetry [%] 8.8 ± 9.2 9.6 ± 9.9 –0.8
(–3.1 to 1.5)
–15.0 to 13.4 0.72
(0.53 to 0.84)
5.1 (55.7) 14.2 (154.5)
Step time variability [s] 0.07 ± 0.04 0.08 ± 0.06 –0.01
(–0.02 to 0.01)
–0.09 to 0.08 0.58
(0.33 to 0.75)
0.03 (42.3) 0.09 (117.2)
Walk ratio [cm/steps/min] 0.45 ± 0.12 0.44 ± 0.12 0.01
(–0.01 to 0.03)
–0.12 to 0.15 0.83
(0.71 to 0.91)
0.04 (10.1) 0.12 (28.1)

SD, standard deviation; CI, confidence interval; LOA, limits of agreement; ICC, intraclass correlation coefficient; SEM, standard error of measurement; MDC, minimal detectable change.