Skip to main content
. 2020 Sep 9;7:138. doi: 10.3389/fnut.2020.00138

Table 2.

Reliability data [Intraclass Correlation Coefficients (ICC), Confidence Intervals (CI), and Typical Error of the Measurement (TEM)] collected during the Pilot Study for pH and sHCO3- measures obtained from the ABL 80 Flex (Radiometer, Copenhagen, DK).

Within-sample reliability ICC (95% CI) TEM (95% CI)
pH (au) 0.76 (95% CI 0.66 to 0.84) 0.02 (95% CI 0.02 to 0.04)
sHCO3- (mmol·L−1) 0.99 (95% CI 0.98 to 0.99) 0.4 (95% CI 0.3 to 0.5)
Week-to-week variability ICC (95% CI) TEM (95% CI)
pH (au) 0.30 (95% CI 0.04 to 0.59) 0.04 (95% CI 0.03 to 0.05)
sHCO3- (mmol·L−1) 0.77 (95% CI 0.59 to 0.89) 1.5 (95% CI 1.2 to 1.9)