Table 2. Severe acute respiratory syndrome coronavirus 2 external quality assessment panel homogeneity tests of triplicate test results of 3 samples using the NucliSENS easyMAG extraction and Allplex nCoV-2019 kits performed at the Asan Medical Center, Seoul, South Korea, as a proficiency test provider, March 23, 2020*.

Sample no. E gene
RdRp gene
N gene
Mean Ct %CV Mean Ct %CV Mean Ct %CV
41 33.82 4.5 34.37 1.6 34.38 0.5
42 25.15 1.4 25.67 1.8 26.67 0.5
43 30.33 1.5 31.34 1.5 31.55 0.7
44 ND ND ND ND ND ND
45 27.52 2.2 28.73 0.3 29.05 0.7
46 24.47 0.7 26.12 0.3 27.05 0.2
47 ND ND ND ND ND ND
48 20.68 2.5 22.49 0.6 23.49 0.3
49 31.07 0.8 32.55 0.6 33.32 0.9
50 27.89 0.9 29.45 0.2 30.35 0.6

*Ct, cycle threshold; EQA, external quality assessment; %CV, percentage coefficient of variation; ND, not detected.